Chat with us, powered by LiveChat
File icon representing laboratory documents and scientific records storage
NATA-accredited calibration services

Guarantee the accuracy of your instruments with NATA-accredited calibration performed by qualified professionals.

File icon representing laboratory documents and scientific records storage
NATA-accredited testing services

Benefit from NATA-accredited testing services that provide independent verification of your equipment’s performance.

Icon representing laboratory features and navigation elements
Bio-Decontamination

Maintain a safe and sterile laboratory environment with expert bio-decontamination services.

Settings icon representing configuration options for laboratory or system controls
Environmental Monitoring Systems

Clean Room and Laboratory environmental monitoring system projects and technical support

Icon representing navigation or features in laboratory or scientific interface
IOQ (Installation Operational Qualification) services

Verify the proper installation and functionality of your equipment with IOQ services.

ser-6.Note icon representing laboratory documentation and scientific records
Commissioning & validation services

Ensure your equipment is properly installed, qualified, and meets performance specifications through commissioning and validation services.

Book a service technician online now

Looking to book a LAF tech service technician for repairs, calibration testing and/or maintenance? Complete our online booking form.

bugbox anaerobic workstations - Laftech
Bugbox

Bugbox Anaerobic Workstation. Baker’s anaerobic workstations are engineered specifically to help microbiologists cope with rising workloads and provide the best primary isolation rates.

uv stick UV-C lamps for disinfecting laftech
UV-STICK Portable Germicidal

UV STICK E2/75H-ST-NX-NT is a commercial grade dual lamp germicidal device with control timer panel and four wheels for easy mobility.

purair advanced ductless fume cabinet
Purair Advanced Ductless Fume Hoods

Our Purair advanced series of ductless fume hoods are a series of high efficiency products designed to protect the user and the environment from hazardous vapors generated on the work surface.

Automated Microbial Monitoring
Lighthouse Automated Microbial Monitoring Systems

LIGHTHOUSE designed the Remote Active Count system to meet the industry need for continuous trouble-free operation and flexible running parameters.

safemate cyto cytotoxic cabinets
Safemate CYTO Cytotoxic Cabinets

BioAir Safemate CYTO Cytotoxic Drug Handling Cabinets LAF Technologies is the exclusive Australian agent for BioAir and their world renowned range of professional cytotoxic cabinets.

Aerosol Photometer - Model PH-4
TEC PH-4 Precision Aerosol Photometer

PH-4 digital photometer is world’s most compact and advanced digital Aerosol Photometer utilised for testing HEPA filters.

Safekeeper evidence drying cabinet
Safekeeper Evidence Drying Cabinets

Safekeeper Forensic Evidence Drying Cabinets Evidence Drying Cabinets for Industry Professionals LAF Technologies is one of Australasia’s leading suppliers of latest generation forensic evidence drying cabinets and other specialist forensic equipment.

Memmert Incubator I Series

Available in both the regular Incubator series with SingleDISPLAY controls and the advanced Incubator Plus range with TwinDISPLAY, they are designed to meet the needs of laboratories seeking accuracy, reproducibility, and compliance with international standards.

Laboratory ovens for scientific testing and precise temperature control by Laftech
Memmert Universal Oven UN

From simple drying processes to complex conditioning procedures, the Memmert UN Series provides stable, repeatable results without the complexity of forced air systems.

Liebherr Laboratory Refrigerators Perfection Range

Modern components and precise controls ensure that the substances and materials are stored optimally.

Memmert WTB Series Water Baths

Whether you are incubating samples, thawing media, or performing routine laboratory procedures, the WTB Series ensures reliable and reproducible results.

OmniaPure XS Basic laboratory equipment used for calibration testing and cleanroom monitoring
OmniaPure xs basic

When you need ultrapure water without the fuss, the OmniaPure xs basic is your ideal solution. This compact, cost-effective system delivers Type I water with exceptional purity, boasting a resistivity of 18.2 MΩ x cm and a low TOC value of < 10 ppb.

Venus 1 laboratory equipment used for calibration testing and cleanroom monitoring
BIOSAFE VENUS® Storage

Safeguard your valuable biological and medical samples with the BIOSAFE VENUS® Sample Storage system available from LAF tech, the leading supplier of lab and scientific equipment in Australia.

Class II A2 vs B1 vs B2 Biological Safety Cabinets: Which One Does Your Lab Actually Need?

Laboratory professional working inside a biological safety cabinet while handling scientific samples in a controlled environment

Choosing the wrong Class II biological safety cabinet can create more than a procurement problem. It can affect operator protection, product protection, exhaust design, maintenance access, room balance and future certification.

Class II biological safety cabinets are built to protect the operator, product and environment through controlled inward airflow, HEPA-filtered downflow and filtered exhaust. The confusion starts because A2, B1 and B2 cabinets can all sit under the “Class II” label, yet they don’t manage recirculated and exhausted air in the same way.

The right choice depends on what your lab handles, how the cabinet will be installed and what your facility’s risk assessment requires.

Start with the work, not the cabinet label

A Class II cabinet is usually considered where biological containment and product protection are both needed. That can include microbiology, biomedical research, pathology, pharmaceutical work, environmental testing and other applications involving biological samples or procedures that may create aerosols.

The cabinet type should follow the work. Ask:

  • Are you handling biological material only, or biological material with small quantities of volatile chemicals?
  • Does the work require product protection, operator protection and environmental protection?
  • Will the cabinet exhaust back into the room or connect to the building exhaust system?
  • Does the room have the air supply and exhaust capacity to support the cabinet?
  • What standard, SOP, regulator expectation or internal quality requirement applies?

A useful warning sign is when a specification asks for “a biosafety cabinet” without stating the application, hazard profile, installation conditions and certification pathway.

For a broader look at cabinet types, testing requirements and selection factors, read our guide: The Complete Guide to Biological Safety Cabinet Testing, Certification & Selection in Australia.

Class II A2 is often the practical choice for biological work

A Class II A2 biological safety cabinet is commonly used for routine microbiological and biomedical applications where protection of the operator, product and environment is required. A2 cabinets recirculate a portion of HEPA-filtered air within the cabinet and discharge HEPA-filtered exhaust either back into the room or through a suitable exhaust connection, depending on the installation and model.

This makes A2 cabinets a practical option for many laboratories because they don’t necessarily require hard ducting. They can also be easier to place within existing facilities, provided the room layout, airflow, heat load, access and certification requirements are checked properly.

Persons working inside biological safety cabinet conducting sterile microbiology research safely

A2 does not mean “suitable for everything.” It is not a substitute for a chemical fume cupboard, and it should not be chosen for work involving hazardous chemical vapours unless the cabinet, risk assessment and facility design support that use.

At LAFtech, we supply and support biological safety cabinets for Australian laboratories that need equipment selection, installation support and ongoing servicing aligned with the way the cabinet will actually be used.

B1 cabinets suit narrower exhaust-driven applications

A B1 biological safety cabinet uses a different airflow pattern. A portion of the air is exhausted through a dedicated external exhaust system, while some air is recirculated through HEPA filtration. This can suit specific procedures where work is performed in the correct region of the cabinet and where the building exhaust system is designed to support the cabinet.

The practical issue is not just cabinet cost. B1 cabinets place more demand on facility design, exhaust connection, airflow verification and user training. Operators need to understand where work should be performed inside the cabinet because airflow zones matter.

A B1 cabinet can be the wrong choice if the lab assumes it behaves like an A2 cabinet with “extra exhaust.” It does not. It needs the right installation context.

Researcher inside biological safety cabinet performing sterile laboratory microbiology procedures

B2 cabinets exhaust all cabinet air

A B2 biological safety cabinet is a total exhaust cabinet. It does not recirculate cabinet air back into the work zone. All inflow and downflow air is exhausted through a dedicated exhaust system after filtration.

This can make B2 cabinets relevant for higher exhaust-demand applications, but it also makes them more demanding to install and operate. They typically require dedicated ducting, suitable building exhaust capacity and careful room air balance. Energy use, make-up air, alarm response, certification access and service planning should all be considered before purchase.

For many labs, B2 is not the safest default choice simply because it exhausts more air. If the facility cannot support the cabinet properly, performance and usability can suffer.

A2 vs B1 vs B2

Cabinet typeAirflow and exhaust patternOften considered forKey planning issue
Class II A2Recirculates some HEPA-filtered air and exhausts some HEPA-filtered airRoutine biological work requiring operator, product and environmental protectionConfirm chemical limitations, room suitability and certification needs
Class II B1Exhausts a defined portion through a dedicated exhaust system and recirculates the balanceSpecific applications needing more direct exhaust in defined work zonesWork position, ducting and building exhaust design matter
Class II B2Exhausts all cabinet air through a dedicated exhaust systemApplications requiring total exhaust cabinet designExhaust capacity, make-up air, energy use and service access

Testing and service history matter after selection

Cabinet choice is only the first step. The cabinet still needs correct installation, certification, servicing and documentation. A cabinet that was suitable on paper can become a risk if airflow is disturbed, filters degrade, alarms are ignored, room conditions change or service records are incomplete.

For regulated or quality-sensitive facilities, documentation matters because it shows that the cabinet has been checked using an appropriate method and within the facility’s quality framework. LAFtech provides NATA-accredited testing services across defined scopes, including testing related to controlled environments and critical equipment. We also support laboratories through servicing, maintenance and technical support where equipment performance needs to be managed over time.

Frequently Asked Questions

Is a Class II A2 cabinet enough for most labs?

For many biological applications, an A2 cabinet is a practical and common choice. The final decision should still account for the material handled, chemical use, room layout, exhaust needs, applicable standards and internal SOPs.

Can a BSC replace a fume cupboard?

No. A biological safety cabinet and a fume cupboard are designed for different hazards. A BSC is not automatically suitable for chemical vapours, even if it has HEPA filtration or external exhaust.

Should procurement choose the highest-exhaust cabinet to be safer?

Not automatically. A B2 cabinet can be the wrong choice if the room and exhaust system cannot support it. Safety depends on matching the cabinet, facility design, use pattern, testing method and maintenance plan.

Choosing between A2, B1 and B2 starts with the real work being performed, not a preference for the most complex cabinet. If your lab is specifying new contamination control equipment or reviewing an existing cabinet, contact our team so we can help align the equipment, testing and support pathway with your facility’s risk profile.